Particle Count Controls

Designed according to the needs of our customers from pharmaceutical industries, BS-Partikel offers a variety of count controls.

PC series

  • Reliable control of accurate particle counting of your device. The amount of particles is 3.000 per mL.
  • Available channel sizes in this moment: 2 µm, 5 µm, 10 µm, 15 µm, 25 µm, 40 µm, 70µm, 100 µm.
  • We also have 25 µm, 40 µm and 70 µm available with 500 particles per mL.

All count controls are delivered in combination with a particle count blank, a certificate of calibration including a record form for an ISO conforming documentation of every performance test.

BS-Partikel offers custom-made confection of various containers of constant particle numbers.

Ph series

  • PharmaCount1510: Mostly used in QC applications in GPM laboratories following directive of <788>. Number of particles and ratio of particles > 10 µm to > 15 µm is set according to the actual LOT of the official USP standard.
  • PharmaCount BC0510: Designed regarding directive <729>. Two populations of particles with nominal mean size of 5 µm and 10 µm. To be (auto)diluted befored measured by light obscuration methods.
  • PharmaCount TC0825: Three populations of particles with nominal mean size of 0.8 µm, 2 µm and 5 µm. To be (auto)diluted befored measured by light obscuration methods.
  • PharmaCount JP: 1.000 Part./mL, nom. mean diameter: 10 µm.

Oi series

  • Suspension of SiO2 (8631 ISO-MTD Dust) in Aeroshell fluid oil for ISO 11171. Volume: 200 mL